Agilent 1200 Infinity Binary Pump (G1312B) (Refurbished)

The Agilent G1312B can be configured for a delay volume as low as 120 µL, making it particularly effective for rapid-gradient HPLC and LC/MS. It combines low system delay with 600 bar pressure capability and precise high-pressure solvent mixing.
Purchase price
5,800
without VAT
In stock
+ 150  delivery

Features

Condition:
Refurbished
Warranty:
3 months
Manufacture year:
0
Countries:
Czech Republic
Additional services
Installation
Extended warranty
Basic training

Features

The Agilent 1200 Infinity Binary Pump G1312B is well suited to fast gradient separations where low delay volume and precise solvent composition are critical. In standard configuration, the pump uses a mixer and pulse damper, while bypassing both reduces pump delay volume to approximately 120 µL. This makes the system particularly useful for fast methods using 2.1 mm ID columns and applications requiring minimal gradient delay.

Key Features

  • High-pressure binary gradient mixing
  • Low Delay Volume down to 120 µL
  • Optimized for fast gradients
  • Up to 600 bar
  • Flow rates up to 5 mL/min
  • Low pressure and composition ripple
  • Multi-point solvent compressibility compensation
  • Automatically optimized piston stroke
  • Suitable for 2.1 mm ID columns
  • Optional 200 µL Low Volume Mixer

Technical Specification

Physical Specifications

  • Weight: 15.5 kg (34 lbs)
  • Dimensions (H x W x D): 180 mm x 345 mm x 435 mm (7 x 13.5 x 17 inches)
  • Power Requirements:
    • Voltage: 100–240 VAC ±10%
    • Frequency: 50 or 60 Hz ±5%
    • Consumption: 220 VA, 74 W / 253 BTU
  • Operating Conditions:
    • Temperature: 4°C to 55°C
    • Humidity: <95% relative humidity (non-condensing)
    • Altitude: Up to 2000 m (operating)

Performance Specifications

  • Flow Rate:
    • Settable Range: 0.001–5 mL/min (in 0.001 mL increments)
    • Accuracy: ±1% or ±10 µL/min, whichever is greater
    • Precision: ≤0.07% RSD or ≤0.02 min SD
  • Pressure Range: 0–600 bar (0–8700 psi)
    • Pressure Pulsation: Typically <1.3% or <0.3 MPa
  • Gradient Formation: High-pressure binary mixing
  • Delay Volume:
    • Standard Configuration: 600–800 µL
    • Low Delay Configuration: 120 µL
  • Composition Control:
    • Range: 0–100%
    • Precision: <0.15% RSD

Connectivity and Control

  • Interfaces: CAN, RS-232C, optional LAN
  • Analog Outputs: Pressure monitoring (1.33 mV/bar)
  • Control Software: Compatible with Agilent ChemStation, EZChrom, OpenLab, and MassHunter platforms