Agilent 1200 Infinity Binary Pump (G1312B) (Refurbished)

The Agilent 1200 Infinity Binary Pump G1312B delivers high-pressure binary gradients at up to 600 bar and 5 mL/min. Its dual-channel architecture provides accurate solvent delivery and reliable performance across routine and high-speed LC methods.
Purchase price
5,800
without VAT
In stock
+ 150  delivery

Features

Condition:
Refurbished
Warranty:
3 months
Manufacture year:
0
Countries:
Czech Republic
Additional services
Installation
Extended warranty
Basic training

Features

The Agilent 1200 Infinity Binary Pump (G1312B) is a high-pressure binary solvent delivery module for HPLC and UHPLC separations. It combines two independent pumping channels with high-pressure solvent mixing and delivers flow rates up to 5 mL/min at pressures up to 600 bar. Its dual-piston-in-series design, active inlet valves, and solvent compressibility compensation provide accurate mobile-phase delivery and stable chromatographic performance.

Technical Specification

Physical Specifications

  • Weight: 15.5 kg (34 lbs)
  • Dimensions (H x W x D): 180 mm x 345 mm x 435 mm (7 x 13.5 x 17 inches)
  • Power Requirements:
    • Voltage: 100–240 VAC ±10%
    • Frequency: 50 or 60 Hz ±5%
    • Consumption: 220 VA, 74 W / 253 BTU
  • Operating Conditions:
    • Temperature: 4°C to 55°C
    • Humidity: <95% relative humidity (non-condensing)
    • Altitude: Up to 2000 m (operating)

Performance Specifications

  • Flow Rate:
    • Settable Range: 0.001–5 mL/min (in 0.001 mL increments)
    • Accuracy: ±1% or ±10 µL/min, whichever is greater
    • Precision: ≤0.07% RSD or ≤0.02 min SD
  • Pressure Range: 0–600 bar (0–8700 psi)
    • Pressure Pulsation: Typically <1.3% or <0.3 MPa
  • Gradient Formation: High-pressure binary mixing
  • Delay Volume:
    • Standard Configuration: 600–800 µL
    • Low Delay Configuration: 120 µL
  • Composition Control:
    • Range: 0–100%
    • Precision: <0.15% RSD

Connectivity and Control

  • Interfaces: CAN, RS-232C, optional LAN
  • Analog Outputs: Pressure monitoring (1.33 mV/bar)
  • Control Software: Compatible with Agilent ChemStation, EZChrom, OpenLab, and MassHunter platforms
  • Local Operation: Agilent Instant Pilot (Revision B.02.00 or higher)