The Agilent Intuvo 9000 GC System with a split/splitless (S/SL) inlet and flame ionization detector (FID) is a modern gas chromatography platform developed for high analytical performance, precise chromatographic control, and simplified maintenance. The Intuvo system uses direct column heating rather than a conventional oven, allowing faster heating and cooling, more efficient analytical workflows, and reduced energy consumption.
The offered system is equipped with a split/splitless (S/SL) inlet with electronic pneumatic control (EPC), a flame ionization detector (FID), and an Agilent J&W DB-1 Intuvo GC column, 30 m × 0.250 mm × 0.25 µm.
Key Features
- Split/splitless (S/SL) inlet with full electronic pneumatic control
- Flame ionization detector (FID)
- Agilent J&W DB-1 Intuvo GC column, 30 m × 0.250 mm × 0.25 µm
- Expandable with TCD, ECD, NPD, FPD, SCD/NCD, or MS
- Click-and-Run technology for rapid column replacement
- Integrated Guard Chip for column protection
- Smart ID Key for automatic column identification
- Color touchscreen with built-in contextual help
- Web interface for remote system monitoring
- Automatic leak checking
- Automatic compensation for ambient pressure and temperature changes
- Pressure control resolution of 0.001 psi
- Compact footprint requiring approximately half the bench space of a conventional GC
Configuration options
- Autosampler Agilent 7693A GC Autosampler G4513A (Second-hand)
- TCD detector or GC/MS Agilent 5975, 5977, 7000, 7010
- Chromatography software Agilent OpenLab or DataApex Clarity
Technical Specifications:
Gas Chromatograph
- Model: Agilent Intuvo 9000 GC
- PN: G3950A
- Direct-heating column architecture
- Retention time repeatability: <0.008% or <0.0008 min
- Peak area repeatability: <1% RSD
Installed Configuration
- Inlet: split/splitless (S/SL)
- Detector: flame ionization detector (FID)
- Column: Agilent J&W DB-1 Intuvo GC
- Column dimensions: 30 m × 0.250 mm × 0.25 µm
- Column PN: 122-1032-INT
Supported Modules
- Inlets: S/SL, MMI
- Other supported detectors: TCD, ECD, NPD, FPD, SCD/NCD
- Compatible with single and triple quadrupole GC/MS
- Supports up to four detector signals
Chromatographic System
- Up to two 30 m × 0.320 mm capillary columns or one 60 m × 0.320 mm column
- Column IDs: 0.150–0.320 mm
- Operating temperature: Ambient +10 °C to 450 °C
- Temperature resolution: 0.1 °C
- Up to 20 temperature ramps and 21 plateaus
- Maximum heating rate: 250 °C/min
- Cool-down from 450 °C to 50 °C in less than 3 minutes
- Maximum run time: 999.99 minutes
Electronic Pneumatic Control (EPC)
- Pressure resolution: 0.001 psi
- Pressure setting increments: 0.001 psi from 0 to 99.999 psi; 0.01 psi from 100 to 150 psi
- Pressure range: 0–100 or 0–150 psig, depending on configuration
- Selectable units: psi, kPa, or bar
- Constant pressure, pressure ramp, constant flow, and flow ramp modes
- Automatic atmospheric pressure and temperature compensation
- Automatic leak checking
- Supports He, H₂, N₂, and Ar/CH₄ carrier gases
Split/Splitless Inlet
- Maximum operating temperature: 400 °C
- Split ratios up to 12,500:1
- Splitless mode with pressure pulsing
- Gas Saver mode
- Electronic septum purge control
- Total flow range:
- 0–500 mL/min (N₂)
- 0–1250 mL/min (He or H₂)
- Turn-Top inlet sealing system
- Optional inert inlet configuration