The Agilent 1200 Series Fraction Collector G1364C is an analytical-scale fraction collection module designed for precise, automated fraction collection at flow rates below 10 mL/min. It supports a wide range of vial, tube, plate, and funnel configurations and integrates seamlessly with Agilent 1200 Series LC detectors and control modules.
Key Features
- Analytical-scale fraction collection for flow rates up to 10 mL/min (higher flows possible with modification)
- Multiple trigger modes: time-based, peak-based, timetable, and manual triggering
- Discrete and continuous flow collection modes
- Low internal delay volumes for high chromatographic fidelity
- Fast-switching low-volume diverter valve (<100 ms switching time)
- Automatic tray and vessel detection
- Optional cooling via G1330B Fraction Collector Thermostat
- GLP support with electronic maintenance and error logs
- Leak detection and exhaust fan for safe handling of hazardous solvents
- CAN-based system integration with optional LAN and RS232 connectivity
Technical Specifications (from Agilent brochure)
Flow & Performance
- Maximum flow rate: 10 mL/min (analytical scale)
- Maximum pressure at diverter valve: 6 bar
- Diverter valve internal volume: ~15 µL
- Switching time: <100 ms
- Delay volumes (typical):
- Inlet to diverter valve: ~50 µL
- Diverter valve to needle: ~10 µL
- Needle: ~4 µL
- Delay calibration sensor: Single-wavelength absorbance detector (654 nm LED)
Fraction Collection Modes
- Trigger modes:
- Time slices
- Peak-based (threshold, up/down slope)
- Timetable (time + peak)
- Manual trigger (with G1323B control module)
- Operating modes:
- Discrete fractions (default)
- Continuous flow (deep well plates only)
- Needle-into-location mode
- Droplet setup mode (bubble-free small fractions)
Fraction Vessels & Capacity
- Supports:
- Well plates (MTP, deep well plates)
- Vials (2 mL)
- Test tubes (12–30 mm OD)
- Funnel trays for external containers
- Maximum tube height:
- 48 mm (standard needle)
- 75 mm (short needle assembly G1364-87202)
- Maximum tube volume:
- ~20 mL (48 mm tubes)
- ~30 mL (75 mm tubes)
- Unlimited with external funnels
- Automatic tray identification
Detector Compatibility
- Fully supported:
- Agilent 1200 Series DAD/MWD (G1315, G1365)
- Fluorescence detector
- LC-MSD (G1946C/D, G1956A/B)
- Other detectors usable but not officially supported for fraction triggering
Interfaces & System Integration
- CAN (Controller Area Network)
- Optional:
- LAN
- RS232C
- External contacts interface
- APG-remote I/O
- Direct interface to G1330A/B Thermostat
- CAN-DC-out for Agilent-approved external devices
Safety & Compliance
- Leak detection with safe leak handling
- Exhaust fan for solvent vapor removal
- Early Maintenance Feedback (EMF)
- Safety Class I instrument
- Certified to IEC, CSA, UL
- Installation Category II, Pollution Degree 2
- Indoor use only
Physical Specifications
- Dimensions (H × W × D): 200 × 345 × 440 mm
- Weight: 13.5 kg
- Power supply: 100–240 VAC, ±10 %
- Frequency: 50/60 Hz
- Power consumption: max. 180 W (active), 200 VA (apparent)
- Operating temperature: 4–55 °C
- Humidity: <95 % (non-condensing)
- Operating altitude: up to 2000 m