Peak Shape Problems in Gas Chromatography

We, 8.10.2025 | Original article from: Phenomenex
Negative peaks in GC can affect your analysis accuracy. Learn about the main causes, from detector overload to contamination, and how to correct them effectively.
<p><strong>Phenomenex:</strong> Peak Shape Problems in Gas Chromatography</p>

Phenomenex: Peak Shape Problems in Gas Chromatography

Negative Peaks - Some or All Peaks Dip Below the Baseline

Negative peaks, where some or all peaks dip below the baseline, can occur in Gas Chromatography (GC) analysis and might impact the accuracy and reliability of your results. In this month's technical tip, we will explore the possible causes and suggest remedies to troubleshoot the presence of negative peaks.

Symptom - Some Peaks Dip Below the Baseline

Phenomenex: Peak Shape Problems: Symptom - Some Peaks Dip Below the BaselinePhenomenex: Peak Shape Problems: Symptom - Some Peaks Dip Below the Baseline

Possible Cause | Suggested Remedy
  • Detector overload in element specific detectors such as ECD, NPD, FPD, etc. can produce both positive and negative peaks: Have the compounds of interest arrive at the detector at a different time from the solvent or other compounds in high concentration. H₂ produces negative peaks with a TCD and helium carrier gas.
  • Dirty ECD detector can give a negative peak after a positive one: Clean or replace the ECD detector.
  • Sample contaminants (hydrocarbons or other non-responders) are present when using ECD, PID, or NPD (thermoionic specific) detectors: Improve sample preparation and cleanup methods prior to injection.

Symptom - All Peaks Dip Below the Baseline

Phenomenex: Peak Shape Problems: Symptom - All Peaks Dip Below the BaselinePhenomenex: Peak Shape Problems: Symptom - All Peaks Dip Below the Baseline

Possible Cause | Suggested Remedy
  • Incorrect polarity of the recorder connections results in nearly all peaks being negative: Reverse polarity of recorder connections.
  • Recorder-integrator wires reversed: Correct connections.
  • Sample injected onto the wrong column for dual-column setups: Reinject the sample onto the correct column.

Symptom - Dip after Solvent Peak

Phenomenex: Peak Shape Problems: Symptom - Dip after Solvent PeakPhenomenex: Peak Shape Problems: Symptom - Dip after Solvent Peak

Possible Cause | Suggested Remedy
  • Detector contamination: Clean or bake out the detector.
  • Sample contamination: For PID detectors, check that the sample has not been contaminated with methanol or water. If necessary, prepare a fresh sample.
  • Often normal for NPD (thermoionic specific) detectors: No correction necessary.

Negative peaks in GC analysis can be addressed by understanding the possible causes and implementing the appropriate remedies. Most of the causes that has been observed are related to the detector that play a crucial role in negative peaks troubleshooting.

Phenomenex
LinkedIn Logo
 

Related articles

Peak Shape Problems: Broad Solvent Peaks/Fronts
Article | Academy

Peak Shape Problems: Broad Solvent Peaks/Fronts

Learn how to troubleshoot and fix broad solvent peaks/fronts in GC analysis with practical tips from Phenomenex to improve peak shape and accuracy.
Peak Shape Problems: No Peaks
Article | Academy

Peak Shape Problems: No Peaks

Missing peaks in GC runs? This technical tip reviews key causes like syringe, inlet, detector, or column issues, with practical solutions to restore results.
Peak Shape Problems: Sensitivity Loss
Article | Academy

Peak Shape Problems: Sensitivity Loss

Learn how to identify and fix sensitivity loss in GC. Discover key causes, from column contamination to injector leaks, and apply remedies for reliable, accurate results.
Phenomenex technical tip for Baseline problems - offset
Article | Academy

Phenomenex technical tip for Baseline problems - offset

Learn how to fix baseline offset in gas chromatography. Identify causes like voltage changes, contamination, or gas flow issues and apply targeted maintenance for stable results.
 

Related articles

Peak Shape Problems: Broad Solvent Peaks/Fronts
Article | Academy

Peak Shape Problems: Broad Solvent Peaks/Fronts

Learn how to troubleshoot and fix broad solvent peaks/fronts in GC analysis with practical tips from Phenomenex to improve peak shape and accuracy.
Peak Shape Problems: No Peaks
Article | Academy

Peak Shape Problems: No Peaks

Missing peaks in GC runs? This technical tip reviews key causes like syringe, inlet, detector, or column issues, with practical solutions to restore results.
Peak Shape Problems: Sensitivity Loss
Article | Academy

Peak Shape Problems: Sensitivity Loss

Learn how to identify and fix sensitivity loss in GC. Discover key causes, from column contamination to injector leaks, and apply remedies for reliable, accurate results.
Phenomenex technical tip for Baseline problems - offset
Article | Academy

Phenomenex technical tip for Baseline problems - offset

Learn how to fix baseline offset in gas chromatography. Identify causes like voltage changes, contamination, or gas flow issues and apply targeted maintenance for stable results.
 

Related articles

Peak Shape Problems: Broad Solvent Peaks/Fronts
Article | Academy

Peak Shape Problems: Broad Solvent Peaks/Fronts

Learn how to troubleshoot and fix broad solvent peaks/fronts in GC analysis with practical tips from Phenomenex to improve peak shape and accuracy.
Peak Shape Problems: No Peaks
Article | Academy

Peak Shape Problems: No Peaks

Missing peaks in GC runs? This technical tip reviews key causes like syringe, inlet, detector, or column issues, with practical solutions to restore results.
Peak Shape Problems: Sensitivity Loss
Article | Academy

Peak Shape Problems: Sensitivity Loss

Learn how to identify and fix sensitivity loss in GC. Discover key causes, from column contamination to injector leaks, and apply remedies for reliable, accurate results.
Phenomenex technical tip for Baseline problems - offset
Article | Academy

Phenomenex technical tip for Baseline problems - offset

Learn how to fix baseline offset in gas chromatography. Identify causes like voltage changes, contamination, or gas flow issues and apply targeted maintenance for stable results.
 

Related articles

Peak Shape Problems: Broad Solvent Peaks/Fronts
Article | Academy

Peak Shape Problems: Broad Solvent Peaks/Fronts

Learn how to troubleshoot and fix broad solvent peaks/fronts in GC analysis with practical tips from Phenomenex to improve peak shape and accuracy.
Peak Shape Problems: No Peaks
Article | Academy

Peak Shape Problems: No Peaks

Missing peaks in GC runs? This technical tip reviews key causes like syringe, inlet, detector, or column issues, with practical solutions to restore results.
Peak Shape Problems: Sensitivity Loss
Article | Academy

Peak Shape Problems: Sensitivity Loss

Learn how to identify and fix sensitivity loss in GC. Discover key causes, from column contamination to injector leaks, and apply remedies for reliable, accurate results.
Phenomenex technical tip for Baseline problems - offset
Article | Academy

Phenomenex technical tip for Baseline problems - offset

Learn how to fix baseline offset in gas chromatography. Identify causes like voltage changes, contamination, or gas flow issues and apply targeted maintenance for stable results.