Peak Shape Problems: Tailing Peaks

We, 6.8.2025 | Original article from: Phenomenex
GC column issues often arise from improper installation. The table below helps you diagnose and resolve common problems using your Column Performance Test Mixture.
<p><strong>Phenomenex: </strong>Peak Shape Problems: Tailing Peaks</p>

Phenomenex: Peak Shape Problems: Tailing Peaks

More often than not, GC column problems are traceable to something improperly done during installation. The table below may assist you in identifying and correcting common installation problems highlighted by your Column Performance Test Mixture. 

Symptom: Tailing Peaks

Possible Cause   /   Suggested Remedy

  • Contaminated or active injector liner or column
    • Clean or replace injector liner.
    • Solvent rinse or replace the column.
  • Dead volume due to poorly installed liner or column: Reinstall liner and column as necessary.
  • Ragged column end: Score the tubing lightly with a sapphire scribe or a ceramic scorer wafer before breaking it. Examine the end (a 20-power magnifier glass is recommended). If the break is not clean and the end square, cut the column again. Point the end down while breaking it, and while installing a nut and ferrule, to prevent fragments from entering the column. Reinstall the column.
  • A bad match between the polarities of the stationary phase and the solvent: Change the solvent.
  • A cold region in the sample flow path: Remove any cold zones in the flow path.
    Debris in the liner or column
    • Clean or replace the liner.
    • Cut a centimeter or two off the ends of the column and reinstall.
  • Injection takes too long: Improve injection technique.
  • Split ratio is too low: Increase split ratio.
  • Overloading the inlet: Decrease the sample volume.
  • Some types of compounds such as alcoholic amines, primary and secondary amines, and carboxylic acids tend to tail
    • Try a different column.
    • Make a derivative of the sample.
Phenomenex
LinkedIn Logo
 

Related articles

Peak Shape Problems: Broad Solvent Peaks/Fronts
Article | Academy

Peak Shape Problems: Broad Solvent Peaks/Fronts

Learn how to troubleshoot and fix broad solvent peaks/fronts in GC analysis with practical tips from Phenomenex to improve peak shape and accuracy.
Peak Shape Problems: No Peaks
Article | Academy

Peak Shape Problems: No Peaks

Missing peaks in GC runs? This technical tip reviews key causes like syringe, inlet, detector, or column issues, with practical solutions to restore results.
Peak Shape Problems: Sensitivity Loss
Article | Academy

Peak Shape Problems: Sensitivity Loss

Learn how to identify and fix sensitivity loss in GC. Discover key causes, from column contamination to injector leaks, and apply remedies for reliable, accurate results.
GC Columns - Troubleshooting & Maintenance
Article | Academy

GC Columns - Troubleshooting & Maintenance

A practical guide to GC column troubleshooting. Learn how to identify and resolve peak tailing, loss of resolution, ghost peaks, baseline drift, sensitivity issues, and retention time shifts.
 

Related articles

Peak Shape Problems: Broad Solvent Peaks/Fronts
Article | Academy

Peak Shape Problems: Broad Solvent Peaks/Fronts

Learn how to troubleshoot and fix broad solvent peaks/fronts in GC analysis with practical tips from Phenomenex to improve peak shape and accuracy.
Peak Shape Problems: No Peaks
Article | Academy

Peak Shape Problems: No Peaks

Missing peaks in GC runs? This technical tip reviews key causes like syringe, inlet, detector, or column issues, with practical solutions to restore results.
Peak Shape Problems: Sensitivity Loss
Article | Academy

Peak Shape Problems: Sensitivity Loss

Learn how to identify and fix sensitivity loss in GC. Discover key causes, from column contamination to injector leaks, and apply remedies for reliable, accurate results.
GC Columns - Troubleshooting & Maintenance
Article | Academy

GC Columns - Troubleshooting & Maintenance

A practical guide to GC column troubleshooting. Learn how to identify and resolve peak tailing, loss of resolution, ghost peaks, baseline drift, sensitivity issues, and retention time shifts.
 

Related articles

Peak Shape Problems: Broad Solvent Peaks/Fronts
Article | Academy

Peak Shape Problems: Broad Solvent Peaks/Fronts

Learn how to troubleshoot and fix broad solvent peaks/fronts in GC analysis with practical tips from Phenomenex to improve peak shape and accuracy.
Peak Shape Problems: No Peaks
Article | Academy

Peak Shape Problems: No Peaks

Missing peaks in GC runs? This technical tip reviews key causes like syringe, inlet, detector, or column issues, with practical solutions to restore results.
Peak Shape Problems: Sensitivity Loss
Article | Academy

Peak Shape Problems: Sensitivity Loss

Learn how to identify and fix sensitivity loss in GC. Discover key causes, from column contamination to injector leaks, and apply remedies for reliable, accurate results.
GC Columns - Troubleshooting & Maintenance
Article | Academy

GC Columns - Troubleshooting & Maintenance

A practical guide to GC column troubleshooting. Learn how to identify and resolve peak tailing, loss of resolution, ghost peaks, baseline drift, sensitivity issues, and retention time shifts.
 

Related articles

Peak Shape Problems: Broad Solvent Peaks/Fronts
Article | Academy

Peak Shape Problems: Broad Solvent Peaks/Fronts

Learn how to troubleshoot and fix broad solvent peaks/fronts in GC analysis with practical tips from Phenomenex to improve peak shape and accuracy.
Peak Shape Problems: No Peaks
Article | Academy

Peak Shape Problems: No Peaks

Missing peaks in GC runs? This technical tip reviews key causes like syringe, inlet, detector, or column issues, with practical solutions to restore results.
Peak Shape Problems: Sensitivity Loss
Article | Academy

Peak Shape Problems: Sensitivity Loss

Learn how to identify and fix sensitivity loss in GC. Discover key causes, from column contamination to injector leaks, and apply remedies for reliable, accurate results.
GC Columns - Troubleshooting & Maintenance
Article | Academy

GC Columns - Troubleshooting & Maintenance

A practical guide to GC column troubleshooting. Learn how to identify and resolve peak tailing, loss of resolution, ghost peaks, baseline drift, sensitivity issues, and retention time shifts.