Agilent ICP-MS Journal (August 2006 – Issue 28)

Ostatní | 2006 | Agilent TechnologiesInstrumentace
Speciační analýza
Zaměření
Průmysl a chemie
Výrobce
Agilent Technologies

Significance of the Topic


ICP-MS has evolved into a versatile platform not only for routine trace-element assays in liquid matrices but also for direct analysis of gaseous and complex samples. Its high sensitivity, coupled with collision/reaction cell technology and automated sample introduction, opens new possibilities in environmental monitoring, industrial quality control, and clinical speciation studies.

Objectives and Overview


This issue presents a series of developments and success stories around Agilent’s 7500 Series ICP-MS instruments and associated accessories. Key topics include:
  • Direct real-time analysis of particulate matter in gas samples via a novel gas-exchanger interface.
  • Certification of an ICP-MS-based multi-element water analysis method at Nestlé Waters’ global quality center.
  • Introduction of ChemStation software revision B.03.03 with enhanced tuning, data-acquisition and batch-review features.
  • Routine speciation of five arsenic compounds in human urine employing an optimized HPLC-ICP-MS procedure.
  • New all-PFA inert sample introduction kit and concentric nebulizer for enhanced long-term stability with challenging matrices.
  • Discussion of collision cell compliance for EPA Method 200.8 in drinking water analysis.

Methodology and Instrumentation


Several novel or enhanced components and workflows are described:
  • Gas-Exchanger System
    • Counter-current argon sweep through a porous tube to capture sub-micron particulates from sample gases (flow up to 500 mL/min).
    • Aspirator and pressure equalizer maintain constant gas delivery into the plasma.
    • Agilent 7500cs ICP-MS for high-sensitivity detection of metal particulates (0.01–0.1 µm).
  • Water Analysis at Nestlé Waters
    • Agilent 7500ce ICP-MS equipped with Octopole Reaction System (ORS) to remove interferences.
    • ISIS autosampler and CETAC ASX-520 for high throughput.
    • Validation under COFRAC accreditation against ISO 17294-2:200 (62-element suite, <1 µg/L detection).
  • ChemStation Rev. B.03.03 Software
    • Multimode autotune (no-gas, He, H₂) in one click; system-wide tuning parameters.
    • Pre-run monitor for up to six masses; offline method editing during runs.
    • Batch View table display of concentration and CPS in real time; graph summary of calibration curves.
  • Arsenic Speciation in Urine
    • Agilent 1100 Series HPLC (isocratic, vacuum degasser, thermostatted compartment) coupled to 7500ce ICP-MS.
    • Agilent G3288-80000 anion-exchange column and G3154-65002 guard column; phosphate/EDTA/acetonate mobile phase at pH 11.
    • MicroMist glass concentric nebulizer; direct 5 µL injection of undiluted urine yields <0.1 µg/L LOD for As species.
  • Inert Sample Introduction Kit and Nebulizer
    • All-PFA kit (G3285A) with o-ring-free torch connector, push-fit design and optional second gas port for O₂ addition.
    • PFA concentric nebulizer (200 µL/min) compatible with inert kit and integrated autosampler probes.
  • Collision Cell Compliance for EPA Method 200.8
    • Discussion of quality-assurance criteria needed to validate CRC-ICP-MS for regulatory drinking water analysis.
    • Support for performance-based interference check standards in lieu of mathematical correction equations.

Main Results and Discussion


The gas-exchanger system maintained full sensitivity for Mo and W reference gases even with up to 500 mL/min of N₂, CO₂ or 10% SO₂ sample flows. Ambient air analysis revealed real-time iron spikes from suspended particulates. At Nestlé Waters, the 7500ce ORS ICP-MS replaced dual OES/AAS workflows, improving detection limits, throughput and operational flexibility under COFRAC accreditation for 56 elements. The ChemStation B.03.03 upgrade reduced manual tuning time, enabled simultaneous multimode autotuning, and provided real-time batch monitoring. In clinical speciation, the new column chemistry fully resolved As(III), As(V), DMA, MMA and arsenobetaine within 12 minutes, eliminating chloride-based ArCl interference. The all-PFA kit and nebulizer deliver droplet-free operation and long-term stability for challenging solvents and high dissolved solids matrices.

Benefits and Practical Applications


  • Direct gas-phase ICP-MS analysis for ambient air monitoring, exhaust gas surveillance and rapid screening of particulate-bound metals.
  • Single-instrument, multi-element water testing with trace-level detection for global beverage quality control and regulatory compliance.
  • Streamlined IQ/OQ and routine operations via advanced software features, reducing downtime and operator workload.
  • Fast, sensitive arsenic speciation in clinical and environmental urine samples without dilution or complex sample prep.
  • Robust, inert sample introduction options for organic solvents, high-matrix samples and long sequence stability.
  • Roadmap toward EPA acceptance of collision cell-enabled ICP-MS for drinking water under Method 200.8.

Future Trends and Opportunities


Ongoing research aims to standardize calibration protocols for metal gas standards and to extend real-time gas monitoring to automotive exhaust and industrial emissions. In beverage and process water, the focus will shift to ultra-trace analytes—such as bromine in added salts and iodine in processing aids—and mercury speciation at sub-ppb levels. Regulatory bodies and accreditation programs are anticipated to finalize CRC-ICP-MS quality-assurance guidelines, broadening approved applications across matrices.

Conclusion


The integration of innovative sample-introduction hardware, advanced collision/reaction cell technology, and user-friendly software substantially enhances the versatility and performance of Agilent’s 7500 Series ICP-MS platforms. These developments support high-sensitivity, high-throughput analyses across gas, liquid and complex matrices, meeting the evolving demands of environmental monitoring, industrial QA/QC, and clinical speciation.

Obsah byl automaticky vytvořen z originálního PDF dokumentu pomocí AI a může obsahovat nepřesnosti.

PDF verze ke stažení a čtení
 

Podobná PDF

Agilent ICP-MS Journal (May 2006 – Issue 27)
Agilent ICP-MS Journal (May 2006 – Issue 27)
2006|Agilent Technologies|Ostatní
Agilent ICP-MS Journal (February 2007 – Issue 30)
Agilent ICP-MS Journal (February 2007 – Issue 30)
2007|Agilent Technologies|Ostatní
Agilent ICP-MS Journal (Issue 32/2007)
Agilent ICP-MS Journal (Issue 32/2007)
2007|Agilent Technologies|Ostatní
Agilent ICP-MS Journal (May 2005 – Issue 23)
Agilent ICP-MS Journal (May 2005 – Issue 23)
2005|Agilent Technologies|Ostatní